MAX4000

Intuitive Interface

Large Comprehensive Display

Digital Filter

Easily Portable

Excellent Sensitivity

Easy to Use

Introduction

The MAX 4000 Electrometer is a superior instrument for accurate measurement of an extremely wide range of all brachytherapy sources, low dose rate, high dose rate, and cardiovascular. It is versatile enough to be used for all external beam and diagnostic x-ray applications.

Brachytherapy

Exceptional sensitivity and a wide dynamic range makes MAX 4000 the electrometer of choice for all brachytherapy measurements. It provides immediate measurements of even the lowest activity isotopes. For example, a 0.27mCi iodine seed, measured in the HDR 1000 Plus Well Chamber, gives a signal of 1.458pA. It can also measure 10Ci and higher iridium sources.

External Beam Radiation Therapy

MAX4000

The MAX 4000 works exceptionally well with all ion chambers typically used for data acquisition in water phantoms, and with all chambers used for quality assurance tests.

Specifications

MAX 4000 ELECTROMETER specifications
Four modes Rate, Charge, Timing, and combined Rate and Charge
  Rate:
  Low Range: 0.001 pA - 950.00 pA, 1 fA resolution
  High Range: 0.001nA - 195.00 nA, 1 pA resolution
  Charge:
  Low Range: 0.01 pC - 999,999 nC, 10 fC resolution
  High Range: 0.01 nC - 999,999 nC, 10 pC resolution
  Timing:
  Selectable time periods for charge collection
  Range: 0-600 seconds
  Increments: 15 seconds
  Resolution: 1 second
  Other mode: "free running"
  Combined:
  The accumulated charge and the current readings are displayed simultaneously.
Range Switching
  User selectable: High or Low
Repeatability
  Short term, � 0.1% � 1 count
  Long term, � 0.2% (maximum change over two years)
  Signal setting time (typ)
    signal < 2 nA - 12 seconds
    signal > 2 nA - 3 seconds
Linearity
  � 0.06% typical for all rate and charge settings
Zero Drift
  < 1fA @STP
Display
  LCD, 2 x 20 with 3/16" high letters for easy viewing
Input
  BNC two plug, triaxial connector
Bias Voltage
  Nominal � 300 Volt bias
Five user settings:
  -300, -150, 0, 150, 300 (VDC)
Accuracy
  � 0.3 Volt
Power
  UL listed wall- mounted power supply, 110 or 220 VAC input
  9VDC@300 mA output Internal battery: 10 hrs/charge
Zeroing
  Automatic zero function, user activated
Output
  Isolated RS-232, Bi-directional, 19,200 baud rate, 8 data bits, no parity, 1 stop bit
Dimensions
  Height: 2.75 inches, 7 cm
  Length: 9 inches, 23 cm
  Width: 8.24 inches, 22.2 cm
  Weight: 3.0 lbs., 1.4 kg
 
© Elimpex-Medizintechnik, Spechtgasse 32, A-2340Mödling, Austria
phone +43-2236-410450
fax +43-2236-860957