Mini CT QC Phantom

76-430

Mini CT QC Phantom
Key Features Accurately evaluates:

Introduction

This highly versatile phantom is designed for routine monitoring of the consistency of all the major parameters of computed tomography (CT) image quality and radiation dose. Its unique, compact design allows for unparalleled portability, easy set up and reliable parameter determinations. It is perfect for use by physicists, technologists and service engineers.

Applications

The disc section consists of a 1 inch thick Lucite® disc with a 6 inch diameter. The six large holes are for the placement of inserts for evaluation of CT number consistency and evaluation of image resolution. The four small holes are for inserting an ion chamber at different locations within the phantom. Lucite inserts are provided to fill the four small holes, when necessary. The disc section is attached to a rectangular acrylic bar containing a thin copper wire embedded along a central groove. This section of the phantom is used to evaluate laser beam alignment and accuracy of slice thickness, slice spacing, slice contiguity, and pilot scan to transverse (longitudinal) scan correspondence. This is achieved by exposing a non-screen film (such as Flex Film Cassettes, listed below) placed underneath the phantom, and making several cuts while the phantom is advanced along the gantry in a pre-programmed manner.


Product Specifications
Dimensions
6 inch Ø, 1 inch thick, with six 1.125 inch through-holes and four 0.50 inch through-holes
Lucite disk
The Lucite disk is attached to the side of the base by two removable nylon, slotted screws
Inserts
Phantom is supplied with seven inserts for 1.125 inch holes; 1 each of: Plastic Water®, bone-equivalent, polystyrene, polycarbonate, polyethylene, nylon, and one acrylic high-contrast resolution insert
Lucite base
11.94 inch long x 1.81 inch wide x 0.69 inch thick, with copper wire (approx. 0.020 inch) fixed into a 0.020 inch deep groove centered on the base
Weight
3 lb (1.36 kg)

Model name Description
 76-430  Mini CT QC Phantom, includes seven inserts
 76-430-5555  Mini CT QC Phantom Kit, includes phantom, seven standard inserts, all seven optional inserts, Teflon-bone semi-ring and carrying case

Related Products
07-800 Series Flex Film Cassettes

Carrying Cases
89-430 Carrying Case

Miscellaneous Accessories
76-430-1000 Low Contrast Resolution Insert designed for determining the CT unit’s ability to detect slight differences in contrast. Two materials with very similar CT numbers are incorporated into the low contrast resolution insert to assess the low contrast detection capability of the unit
76-430-1212 Acrylic Insert with Wire, 0.50 inch
76-430-2000 and 76-430-3000 Teflon® and Lung Inserts These inserts provide the CT number and density that are important when treatment planning parameters are being established for radiation therapy patients
76-430-4000 Teflon-Bone Semi-Ring is used as a beam hardening ring for simulating clinical conditions. The ring has been machined to slide easily over the phantom, so that each of the inserts will the effect of beam hardening
76-430-6000 Acrylic Insert
76-430-7000 Fillable Insert
76-430-8000 Aluminum Insert