Introduction
This highly versatile phantom is designed for routine monitoring of the consistency of all the major parameters of computed tomography (CT) image quality and radiation dose. Its unique, compact design allows for unparalleled portability, easy set up and reliable parameter determinations. It is perfect for use by physicists, technologists and service engineers.
Applications
The disc section consists of a 1 inch thick Lucite® disc with a 6 inch diameter. The six large holes are for the placement of inserts for evaluation of CT number consistency and evaluation of image resolution. The four small holes are for inserting an ion chamber at different locations within the phantom. Lucite inserts are provided to fill the four small holes, when necessary. The disc section is attached to a rectangular acrylic bar containing a thin copper wire embedded along a central groove. This section of the phantom is used to evaluate laser beam alignment and accuracy of slice thickness, slice spacing, slice contiguity, and pilot scan to transverse (longitudinal) scan correspondence. This is achieved by exposing a non-screen film (such as Flex Film Cassettes, listed below) placed underneath the phantom, and making several cuts while the phantom is advanced along the gantry in a pre-programmed manner.