The QUART dido2100(K) is especially designed as diagnostic dosemeter for Quality Assurance and Service in any x-ray application. It is the first instrument of its kind which measures the dose-width pro- duct (DWP) with a flat detector based on semiconductor technology. It thus provides access to an elementary parameter to assess either true dose or dose-area product (DAP) at dental panoramic x-ray installations. In addition, to check on the installation’s spine area compensation, the dosimeter evaluates the dose-width product rate after the semi circle. Due to the improved detector design and its highly effective shielding, a maximum precision is achieved for HVL measurements in mammography and R&F applications. The sensitivity of the device ensures precise results for measurements of very low dose rates in general. For precise kV measurements the QUART dido2100K records maximum kV as well as effective kV for every exposure. Due to its technical design the dido can also be used in Digital Subtrac- tion Angiography (DSA) as well as in Radiography and Fluoroscopy (R+F). It provides efficient measurement technology for service experts, x- ray scientists and manufacturers of radiological equipment. |
![]() | QUART dido2100K QUART dido2100 | fully equipped dosemeter version diagnostic dosemeter without kV feature (upgradeable) |
![]() |
The eQA Option For enhanced Quality Assurance (eQA) the QUART dido2100K can be hooked up to a computer via its USB interface. This enables users to evaluate acquired waveforms and to direct-print test protocols. The optional activation for that kind of data output has to be acquired to be able to use this feature. The eQA features of the QUART dido2100(K), however, also include measurement capabilities which similar systems do not provide, such as dose-width product and IEC compliant measurement of irradiation time (already available in any expansion stage). |
| Included in the delivery of your QUART dido2100K: | ||
![]() |
QUART dido2000K | diagnostic dosemeter with transport and storage case, manual and guide book on dose values | QUART cu08 | added filtration for IEC/DIN compliant dose and kV checks |
| QUART al60 | added filtration for QA without a dental phantom (optional) | |
| QUART MK dent | tool for easy detector positioning at dental OPGs (optional) | |
| QUART al250 | added filtration for R+F QA (optional, if required) | |
| Also available: | ||
![]() |
QUART dent/digitest | test phantom for digital dental QA | QUART dent II + mpc | test phantom + holder for conventional dental QA |
| QUART DSA | improved DSA phantom (Type C)) | |
| QUART SP vario | test phantom for enhanced conventional R+F QA | |
| QUART SP digi | test phantom for digital R+F QA | |
| QUART SP_dl | test phantom for digital fluoroscopy | |
| Dose (Width Product) | 0.005 µGy – 999 Gy (*cm2) | |
| Resolution | 0.01 nGy (*cm2) | |
| Range of Application | both attenuated and open radiation beams All radiation qualities are automatically compensated over the full range – for a maximum in measurement precision. |
|
| Inaccuracy | ±5% | |
| Reproducibility | ± 99.5 % | |
| Dose Rate | 0.1 µGy/s – 1 Gy/s | |
| Resolution | 0.1 nGy/s | |
| Quantities | real-time display of dose rate values after first half of exposure (Central) after full exposure (Peak) |
|
| Inaccuracy | ±5% | |
| Reproducibility | ± 99.5 % | |
| Voltage | 25 – 35 (Mammo) & 40 – 150 kV (Diagnostic) | |
| Resolution | 0.1 kV | |
| Range of Application | open radiation beam
Calibrated for determination of tube filtrations featuring 2.5 mm Al (diagnostic) or 30 µm Mo (Mammo) potential. attenuated radiation beam Calibrated for kV measurements at tubes featuring 25 mm Al (R&F), 2 mm Al (Mammo) added filtration or external 0.8 mm kV-filter (included in delivery). Correction factors are given for varying radiation qualities. |
|
| Quantities | kVp & effective kV | |
| Inaccuracy | ± 5 % subject to tube potential | |
| Time | 0.5 ms – 20 s | |
| Resolution | 0.1 ms | |
| Quantities | Exposure Time – for full exposure Irradiation Time – Radiation above 50% dose rate level (in accordance with IEC 60601-2-54) |
|
| Inaccuracy | ± 0.5 ms or 0.1 % from displayed value | |
| Pulses | 1 – 65.000 | |
| Resolution | Single Pulse | |
| Trigger Level | 0.1 nGy/s | |
| Inaccuracy | ± 1 Pulse | |
| Data Communication | ||
| USB 1.1 Interface (USB 2.0 compatible) |
||
| data read-out optional (SW II option) | ||
| General | ||
| Single Exposure Method: Only one exposure is necessary to acquire all parameters for QA testing and Service operations! |
||
| Configuration: Auto-Start, Auto-Stop & Auto-Reset | ||
| Power Supply: 9V Alcaline | ||
| Power Consumption <7 mA | ||
| Battery Life: 2 Years when used regularly | ||
| Base Unit: 16.1x7.0x4.4cm (LxWxH) Weight: 235 g (incl. battery) |
||
| Detector: 5.0x1.6x0.4cm (LxWxH) Weight: negligeable (!) |
||
| Target/Filter Range | ||
| W/Al 40 - 150 kV; 2.5 mm Al | ||
| Mo/Mo 25 – 35 kV; 30 µm Mo | ||
| Anode/Filter Combinations | ||
| Mo/Mo (Standard); Mo/Rh; Rh/Al; | ||
| Rh/Rh; W/Rh; more upon request | ||