ADVANCED THERAPY DOSIMETER
MODEL 35040


SPECIFICATIONS

DOSE AND RATE DISPLAY RANGES

Charge Full Scale

Charge Sensitivity

Current* 
Full Scale

Current* 
Sensitivity

200.00 pC

0.01 pC

200.0 pA

0.1 pA

2.0000 nC

0.0001 nC

2.000 nA

0.001 nA

20.000 nC

0.001 nC

20.00 nA

0.01 nA

200.00 nC

0.01 nC

100.0 nA

0.1 nA

2.0000mC

0.0001mC

2.400mC

0.001mC

*Average current is displayed with ten times greater resolution.

EFFECTIVE EXPOSURE TIME RANGES

Full Scale Range

Display Resolution

59.99s

0.01s

59min, 59.9s

0.1s

STABILITY
  • Designed for ultra long-term stability error of approximately 0.1% per five years. 
LEAKAGE
  • Virtually removes effects of total system leakage during measurement. Uncorrected leakage <10fA over temperature range. 
LINEARITY
  • Maximum non-linearity variation from straight line of 0.1% of all charge and current ranges. 
RESOLUTION
  • Resolution 0.005% of range (4-1/2 digits) for charge, dose, and average rate and average current; 0.05% of range (3-1/2 digits) for current and rate. 
WARM-UP
  • Fully meets specifications within five minutes of applying power. 
MEASUREMENT ACCURACY
  • 18°C to 28°C. 
  • Charge ±(0.20% reading plus two counts). 
  • Current ±(0.20% reading plus two counts). 
BIAS
  • Eleven user-programmable steps from -500V to +500V in 0.1-volt increments. 
  • Accuracy ±0.3 volt for loads <0.2mA. 
  • Front panel selectable. 
ION CHAMBER CALIBRATION FACTORS
  • Thirty-two user-programmable calibration factors. 
  • Front panel selectable. 
DISPLAY UNITS
  • All practical radiation and electrical units. 
RS-232 COMPUTER CONFIGURATION
  • For customizing and data transfer. 

 

 


Keithley's Advanced Therapy Dosimeter | Features
 